Portable gas analyzers are microprocessor–controlled instrument with sophisticated software process that meets the needs of industry in analyzing the Gases with accuracy and reliability. Portable Bio Gas Analyzers NDIR transducer provides highly stable and reliable results. We are supplying gas analyzers in India wise.
Portable Bio Gas Analyzer
Vasthi Portable Bio Gas Analyzers is a light weight, easy to handle,battery-powered analyzer, used to verify measurements and for spot checking when other methods provide questionable results.
Key Features
NDIR absorption principle
The large LCD screen with clear display assures simple operation.
Single-beam infrared gas analyzer having excellent stability is adopted.
The O2 sensor is available in 2 types:
Paramagnetic type and
Electro Chemical type.
Automatic in operation
Micro controller technology
Customer specific configurable
Suitable for Single or Multiple GAS Analysis
Long battery backup hours
Alarm indicators-low, high and fault Self diagnosis
Easy calibration in fresh air
Run-while charging capability
Virtually maintenance free
No interference from glycol, methanol or amine contaminants(vapor phase)
Fast and accurate real time measurements
Optional RS 232 and Computer Interface to transfer the shareddata
Data logger to store the data in system,Internet facility to monitor 24 hours
TECHNICAL SPECIFICATIONS
Measured component/ measurement range
CO2 : 0- 200 ppm or 0-100%
CO : 0- 200 ppm or 0-100%
NO : 0-2000 ppm or 0-100%
H2S : 0-5000 ppm or 0-100%
CH4 : 0-2000 ppm or 0-100%
Propane : 0-500 ppm or 0-100%
O2 Electrochemical : 0-10 ppm or 0- 25%
O2 Paramagnetic : 1 % 0-100%
Application Data
Environmental Temperature Range : -20° to 50° C (-4° to 122° F)
Sample Inlet Pressure :10 to 25 PSIA, 10 PSIG maximum
Sample Flow Rate : 0.1-10 L/min (0.2 to 20 scfh)
Electrical & Communications
Input Voltage : 100-240 VAC, 50-60 HZ standard
Electrical Storage :12-Volt, Sealed Lead-Acid Battery Approx. 8 hours usage time per charge
Current :0.5 A @ 120 VAC during recharging
Outputs : RS232 – all parameters
LCD Display :Concentration
Alarms: Two Levels (selectable range)
Accuracy :+2%
Repeatability:+0.5%
Signal response time:t 90 <2sec. (At 1 LPM) 90
Operating Temperature : 0 – 50 Deg. C
Warm up Time :‹3 min